This article explains how to reset the measuring cell if standard methods of clearing a constant 1000ppm reading are unsuccessful.
Checks to try before resetting the cell:
If the above methods are unsuccessful in clearing the 1000ppm fault, please proceed with the following steps:
- Disconnect J26 from the rear of the ODME Computer module and ensure that the valves cannot be opened during this process.
- Ensure that all supplied are available to the ODME system.
- In the “General Settings” menu, record all settings.
- Change the following settings to the values stated below:
Flow Meter 1: Manual
Manual Cap1: 100 m3/hr
Ship Speed: Manual
Man. Speed: 10 kn
Cleanpr.Min: 0.0 Bar
Cleanpr.Max: 6.0 Bar
Workpr.Min: 0.0 Bar
Workpr.Max: 2.0 Bar
- Save and exit.
- Once the settings are saved, enter the “Oil Discharge” mode and enter the following oil discharge settings:
Flowmeter: Cargo
Oil type: 9
Oildisc acc: 0 L
Alarm limit: 100 L
Date: dd:mm:yy
Time: hh:mm:ss
Warning 0 L/nm
- Start Oil Discharge
- Once “Start oil discharge” is selected, a preparation cycle will run. Once the preparation cycle is complete and the discharge screen is displayed, stop the discharge by pressing the enter button.
- Reset all settings to the original values recorded in step 2 and reconnect J26 at the rear of the ODME computer module.
- Retest the system.
This procedure is a "Last resort". If this procedure does not clear the 1000ppm reading, the measuring cell will need replacing.
Remotesupport@rivertrace.com would be extremely grateful on hearing of the results and feedback from this procedure.